Single Point Articulation Test (SPAT)
| 1238888 | 
Mounting should always be tested prior to a SPAT test. The SPAT test checks the repeatability of the 
q	Set up the FARO  compensation cone, or a hole on your part, approximately ⅔ of the 
q In CAM2 2024, start the SPAT routine from the FaroArm Manager Probe Management dialog box.
- 
                                In the FaroArm Manager, click Probe Management. 
- 
                                In CAM2 2024 or CAM2 Measure, click Devices > Device Center, and in the FARO Device Center, click Manage. 
q	In the Probe Management dialog box, click Single Point Articulation Test (SPAT). 
q Place and hold the calibrated probe in the compensation cone or any conical socket.
q	Start with the elbow of the 
q Watch the screen and slowly sweep the elbow so that it ends up on the opposite side.
Measurements automatically record as the elbow moves.
                     
 
Figure 7-8 Single Point Articulating Test (SPAT)
- 
                                In the dialog box, watch the status bar. Sweep the elbow all the way to the opposite side. 
q Watch the screen and slowly sweep the elbow back to the starting position.
7-Axis 
q The SPAT automatically completes. The set of measurements from each sweep are compared for repeatability. If the sets are not repeatable, you must repeat the test.
The 
q	Notice the deviations in X, Y, and Z. The (Max-Min)/2 Error, or Single Point repeatability value, of each coordinate should be lower than the specification values of your 
q Click:
- 
                                Report to generate a detailed report of the test. 
- 
                                Restart or discard the results and repeat the test. 
- 
                                Close button to end the test. 