Gage Calibration

1238888

The FaroArm Manager software has commands that certify the accuracy and repeatability of your Gage. The Single Point Repeatability Test (SPAT) allows you to test the Gage using a hole. Use measuring software such as CAM2, SmartInspect, or CAM2 Measure to test the volumetric accuracy.

The values measured in all of these calibrations include 100% of the measured points. We will always refer to this as the “Max. minus Min. over Two,” or the Maximum measurement subtracting from the Minimum measurement divided by two (Max.-Min.)/2.

If any test fails, print the results and repeat the calibration two more times. Print all of the results and contact FARO ’s Customer Service by Phone, Fax or Technical Support.

Repeatability

Perform the Single Point Repeatability Test (SPAT) to check repeatability of the Gage using a hole or a cone. You must use a firmly secured object with a hole at least 1 mm (0.04 in) smaller than the ball probe. If a cone hole is used, extreme care must be taken to ensure that the probe is seated at each “measurement.”

Volumetric Accuracy

Perform a Sphere to Sphere, or Point to Point length measurement in order to certify the volumetric accuracy (EUni) of the Gage. This should be done using traceable length artifacts, which are measured at various locations and orientations throughout the working volume of the Gage. This test is a method for determining articulating measurement machine accuracy.

See Product Specifications for the specification values (SPAT and EUni) of your Gage.

For optimum performance, correctly mounting and operating the Gage is very important. See Gage Performance Verification Checklist.